Troubleshooting Guide
Crosshead will not stop
•
Servo is improperly connected.
•
Improper machine parameters. Contact an MTS Applications Engineer for
further assistance.
•
Controller error. Contact an MTS Applications Engineer for further
assistance.
Limit is always active —
cannot clear
•
Proximity limit switch on the crosshead is obstructed.
•
Limit switch is defective. Contact an MTS Applications Engineer for further
assistance.
•
Frame controller limit switch circuitry is malfunctioning. Contact an MTS
Applications Engineer for further assistance.
Limit switches not
functioning
•
Defective limit switch. Same as above. DANGER — DO NOT RUN
MACHINE! Contact an MTS Applications Engineer for further assistance.
Will not pull rated
maximum load
•
Defective servocontroller. Contact an MTS Applications Engineer for
further assistance.
•
Load cell is not calibrated.
Improper machine
parameters
•
Contact an MTS Applications Engineer for further assistance.
Load cell will not
calibrate
•
Load cell is not connected.
•
Defective load cell.
•
You are attempting to automatically calibrate a non-MTS load cell. These
must be manually calibrated.
•
Improper calibration information being used. Check the load cell
information in the testing software.
Load cell amplifier
defective
•
Contact an MTS Applications Engineer for further assistance.
Load plateaus
•
Defective load cell.
Unusual noise
•
Grinding sound indicating that a moving part is rubbing. Lubrication is
needed. Inspect belt and screw bearings. Contact an MTS Applications
Engineer for further assistance.
•
Debris caught in the cooling fan.