beautypg.com

J6 and j7 monitor, J8 and j9 dc conditioner – MTS ReNew Technical Reference User Manual

Page 31

background image

ReNew Technical Reference

Installation

31

J6 and J7

Monitor

Two Monitor connectors are provided. There are several possible uses for analog
monitor outputs: external data acquisition, tuning, troubleshooting, etc. For
tuning, it is desirable to monitor command and feedback, or command and error,
simultaneously while changing the controller parameters. Therefore, two monitor
outputs are provided. Specifics for these connectors are:

Analog +/-10.5 V

Calibrated to +/-10 V

16-bit resolution minimum

BNC connectors

Note

The load cell is mounted to the crosshead and its wiring is internal to the
frame, thus its connector is not on the rear panel.

J8 and J9 DC

Conditioner

Two DC Conditioner connectors are provided. The two application specific
transducers might be biaxial strain gage base extensometers. With external
completion resistors, the DC conditioners could be used with quarter bridge
strain gages. Pin assignment is as follows:

Pin

Signal

1

TEDS data

2

EX+

3

EX-

4

FB-

5

RCAL1 (FBR+)

6

RCAL2 (FBR-)

7

FB+

8

EXS-

9

EXS+

10

TEDS ground