Handset Control

Operation

Travel limit switches

(Physical limits)

The manner in which limit switch adjustments are made varies from system to
system depending on the manufacturer. Be sure you understand how to set the
limits on your existing load frame. Your MTS representative will be able to show
you how to set the limits.

Always adjust the limits whenever you change grips or fixtures. Damage to
equipment or personal injury may result.

Always adjust the limits whenever you change grips or fixtures. Damage to
equipment or personal injury can result.

When a physical limit is reached there are three ways to get the crosshead
moving again:

Utilize the crosshead capabilities of your software (virtual handset). Move
the crosshead away from the limit until the switch closes and the crosshead
can move in both directions again. See the TestWorks software manual for
further details.

Manually move the adjustable limit along the guide column away from the
crosshead until the limit switch is no longer active.

Use the manual handset control to move the crosshead until the limit switch
is no longer active.

Load cell mounting

MTS offers a wide range of precision, TEDS tension/compression load cells.
TEDS load cells are automatically recognized by the software, enabling the user
to switch without manual software input. The following figure shows the
antirotation load cell mounting with grip adapter. Please contact your Sales
Engineer for assistance in selecting the proper load cell.

CAUTION