Maintenance and Servicing
8.1.4 Cooling compressor
In order to guarantee perfect
function and long lifetime of the
refrigeration unit, it is absolutely
essential to remove dust deposits
from the condenser ( Fig. 33 , no. 1)
(with a vacuum cleaner, paintbrush
or bottle brush, depending on the
amount).
After the screws (2) have been
loosened, the air inlet cover on the
front (3) can be removed and the
condenser can be cleaned.
After the screws have been loos-
ened, the air inlet cover on the
rear (4) can be removed as well,
to clean the refrigeration module
from both sides if necessary.
8.2 Regular maintenance
Every three months
Every three months
►
Check that the door seals fit tightly and adjust the door if necessary (see page 68 ).
►
In permanent mode: Grease the moving parts of the doors (hinges and lock) with thin
silicon grease and check that the hinge screws are not loose.
Annually
Annually
►
Grease the moving parts of the doors (hinges and lock) with thin silicon grease and check
that the hinge screws are not loose.
►
Check that door seals fit tightly and adjust the door if necessary (see page 68 ).
►
Check safety labels (warning labels on the door) (see page 8 ). Replace them if they are
unrecognisable. They can be ordered from the Memmert customer service.
►
Check and replace the sterile filter if it is dirty (only models with CO
2
supply, see service
manual)
Additional maintenance every two years, for models with CO
Additional maintenance every two years, for models with CO
2
supply
supply
►
Replace all sterile filters (see service manual)
1
2
3
4
Cooling compressor