Memmert ICH 256 Climate Chambers (Generation 2003) User Manual
Page 43

43
Operation and Control
14. Setting the fan speed of the second ramp segment:
Turn the push-turn control to the right until the fan
display flashes.
Hold down the SET key and set the desired fan speed of
50 % (five bars lit up) with the push-turn control.
15. Setting the relative humidity of the second ramp seg-
ment:
Turn the push-turn control to the right until the humid-
ity display flashes.
Hold down the SET key and set the desired humidity
setpoint to 70.0 % rh with the push-turn control.
rh
%
16. Setting the CO
2
content of the second ramp segment
(only for appliances with CO
2
supply, otherwise please
continue with point 17):
Turn the push-turn control to the right until the CO
2
display flashes.
Hold down the SET key and set the desired CO
2
setpoint
to 15.0% with the push-turn control.
CO
2
17. Set lighting of the second ramp segment (only for mod-
els with lighting, otherwise please continue with point
18):
Turn the push-turn control to the left until the lamp uv
display appears.
Hold down the SET key and choose on with the push-
turn control.
Turn the push-turn control to the left until the lamp dl
display appears.
Hold down the SET key and set to On with the push-turn
control.
18. Setting the close statement of the second ramp seg-
ment:
Turn the push-turn control to the right until a segment
close statement, e.g. end, appears.
Hold down the SET key and set the close statement next
with the push/turn control.
19. Setting the duration of the third ramp segment:
Select the time display with the push-turn control.
Hold down the SET key and set the time to 1.00 with the
push-turn control.
h
20. Setting the temperature of the third ramp segment:
Turn the push-turn control to the right until the tem-
perature display flashes.
Hold down the SET key and set to 20.0 °C with the push-
turn control.
°C