Advanced Functions

Setting

Setting

1. Set the desired CO

2

balance point in the SETUP (see page 53 ) and set the accompanying

compensation correction value to 0.0 %.

2. With a reference instrument, measure the deviation in the stationary state in the selected

CO

2

balance point.

3. Under SETUP, adjust the compensation correction value. If the measured reference CO

2

content is too low, the compensation correction value must be set with a negative sign.

4. Perform a control measurement with the reference instrument.
5. The procedure can be performed for the CO

2

balance points 5 %, 10 % and 15 %.

Example: CO

2

deviation at 10 % should be corrected:

1. Set the CO

2

balance point in the SETUP to CO2 10 and set the accompanying compensation

correction value to 0.0 %:

PRINT

SETUP

loop

t3

t4

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

3

4

2

1

STERI

DEFRO

°C

°C

rh

mb

%

CO

mb

2

IN 1

IN 2

OUT

IN 1

IN 2

OUT

MIN

AUTO

MAX

2. With a calibrated reference instrument and at a defined setpoint CO

2

content of 10.0 %, an

actual CO

2

content of 11.5 % is measured.

3. Set the compensation correction value in the SETUP for CO2 10 to 1.5 %:

PRINT

SETUP

loop

t3

t4

t2

t1

on

off

Mo

Tu

We

Th

Fr

Sa

Su

3

4

2

1

STERI

DEFRO

°C

°C

rh

mb

%

CO

mb

2

IN 1

IN 2

OUT

IN 1

IN 2

OUT

MIN

AUTO

MAX

4. The reference instrument should display 10 % after the calibration procedure.
With CO2 5 and CO2 15, further calibrations can be programmed at 5 % and 15 %.

If all compensation correction values are set to 0.0 %, the factory CO

2

calibration

settings are restored.