Operation and Control
Setting the relative humidity of the third ramp segment:
Turn the push-turn control to the right until the humidity
display flashes.
Hold down the SET key and set the desired humidity
setpoint to 50.0 % rh with the push-turn control.
%
rh
22. Setting the CO
2
content of the second ramp segment
(only for appliances with CO
2
supply, otherwise please
continue with point 23):
Turn the push-turn control to the right until the CO
2
display flashes.
Hold down the SET key and set the desired CO
2
setpoint
to 0.0% with the push-turn control.
CO
2
23. Setting the lighting of the second ramp segment (only
for models with lighting, otherwise please continue with
point 24):
Turn the push-turn control to the left until the lamp dl
display appears.
Hold down the SET key and set to on with the push-turn
control.
24. Setting the close statement of the third ramp segment:
Turn push-turn control to the right until a close state-
ment appears in the display, e.g. end, and press the SET
key briefly to confirm.
25. Leave programming write mode EDIT
Turn push-turn control to the right until EXIT appears in
EXIT
EXIT
the display, and press the SET key briefly to confirm.
26. Setting the temperature monitoring:
Turn the push-turn control to the right and adjust the
temperature monitoring (for more detailed information
on this, see page 53 ).
°C
MIN
AUTO
MAX
27. Activating the programme:
Turn the push-turn control to the right until the stop
icon flashes.
flashes.
Hold down the SET key and select the start icon with
with
the push-turn control. Release the SET key, and the pro-
gramme is activated.
5.10.8 Operation with PC/laptop (optional)
The cooled incubator can optionally be used, controlled and programmed with a PC/laptop.
For this purpose, it is equipped with a USB or ethernet interface on the rear side (see page 14
and 57 ).
The control of the appliance with the Memmert computer software “Celsius“ is
described in a separate manual.