Warning Messages and Malfunctions
6.3 Power failure
In case of a power failure, the cooled incubator operates as follows:
In normal
and week time switch operating modes
and week time switch operating modes
After the power supply has been restored, operation is continued with the parameters set.
The time and duration of the power failure are documented in the log memory.
In Programming mode
In Programming mode
►
After a power failure of less than 60 minutes, the current programme is continued from
the point at which it was interrupted. The time and duration of the power failure are
documented in the log memory (see page 58 ).
►
In case of a power failure of more than 60 minutes, the cooled incubator starts in manual
operating mode for safety reasons and all setpoint values are changed to safe default
values (see table below).
For remote operation
For remote operation
If there is a power failure in remote operation, the appliance starts in the normal operating
mode for safety reasons and all setpoint values are changed to safe default values (see table).
The programme can only be continued from the computer. The time and duration of the
power failure are documented in the log memory.
Parameters
Default value
Temperature
20 °C
Fan speed
100 %
CO
2
0 %
rh
off