7 settings example programme mode – Memmert ICH 256 Climate Chambers (Generation 2003) User Manual
Page 40

40
Operation and Control
5.10.7 Settings example programme mode
On Monday at 8.00, the cooled incubator should heat up to 37 °C as quickly as possible with
a fan speed of 30 %, and reach a relative humidity of 70 % rh. For models with CO
2
supply,
the CO
2
content should be set to 5 % and for models with interior lighting, daylight should be
simulated.
Once the temperature and humidity have been reached, the cooled incubator should retain
the setpoint values at a fan speed of 50 % for 45 minutes. For models with lighting, the test
samples should be subject to UV light and daylight during that time.
Afterwards, the appliance should cool down to 20 °C and reach 50 % rh humidity within 1
hour at 30 % fan speed. For models with CO
2
supply, the CO
2
content should be adapted to
the natural content of the ambient air and for models with interior lighting, daylight should
be simulated. The overtemperature protection should trigger at 38.5 °C.
Mo 8.00 h
50 °C
37 °C
20 °C
50 %
30 %
80 %
0.45 h
1.00 h
t
0.01 h
70 % rh
50 % rh
10 %
15 %
5 %
Ramp 1
Ramp 2
Ramp 3
overtemperature limit
DL
UV
no
combination
possible
Fig. 20
Settings example programme mode