Leica Biosystems ANGLE TWO User Manual
Page 35

Alignment
Leica Angle Two
™
User Manual Rev A02 © Leica Biosystems Richmond Inc. 2010
35
5. Attach the clamp adaptor to the dial indicator by sliding the raised area of the dial indicator
into the groove of the adaptor.
Figure 36. The raised area of the dial indicator
slides into the groove of the clamp adaptor
6. Slide the clamp covering towards the middle of the clamp to reveal the clamp hole. Push the
top of the clamp adaptor into the clamp hole and release the covering to lock the adaptor in
position.
Figure 37. Slide the clamp covering to the middle to
insert the clamp adaptor inside
7.
Attach the clamp to the bottom of the alignment shaft and tighten the clamp screw so that it
cannot move.
Clamp adaptor
Dial indicator
Clamp covering
Clamp screw
See also other documents in the category Leica Biosystems Equipment:
- SM2010 R (46 pages)
- CM305 S (66 pages)
- CE Knife holder (14 pages)
- CM1950 (78 pages)
- VT1200/VT1200 S (62 pages)
- VT1000 S (42 pages)
- HI1220 (18 pages)
- HI1210 (18 pages)
- CM1850 UV (56 pages)
- CM1850 UV (54 pages)
- CM1510-S (56 pages)
- CM3600 XP (114 pages)
- CM1100 (36 pages)
- CM1860/CM1860 UV (98 pages)
- Premium blade holder (14 pages)
- CM1520 (92 pages)
- VT1000 A (49 pages)
- VT1000 P (48 pages)
- SP2600 (51 pages)
- SM2500 (76 pages)
- SM2400 (8 pages)
- SM2000R (22 pages)
- EG1160 (32 pages)
- EG1150 C (20 pages)
- EG1150 H (42 pages)
- EG1130 (2 pages)
- EG1120 (2 pages)
- EG F (18 pages)
- ASP300 (70 pages)
- PELORIS_PELORIS II (160 pages)
- TP1020 (102 pages)
- ASP200 S (88 pages)
- ASP300 S (88 pages)
- ASP6025 (178 pages)
- LN22 (44 pages)
- RM CoolClamp (20 pages)
- RM2265 (88 pages)
- RM2255 (82 pages)
- RM2245 (78 pages)
- RM2235 (78 pages)
- RM2125 RTS (60 pages)
- SP9000 (30 pages)
- SP1600 (10 pages)
- ST4040 (50 pages)
- ST5020 (88 pages)