6.5.2 Selecting alarm volume

Press the ‘Immersion time’ button and hold.

Press ‘+’ or ‘-’ respectively:

Each time when pressing a button, the volume changes audibly:

Alarm OFF --> Alarm low --> Alarm loud.

If ‘Alarm OFF’ is selected (not recommended!), the instrument must be
observed constantly, to make sure all slide racks are removed immedi-
ately when reaching the last station!

Proceed to --> 6.6.1 ‘Standby mode’.

6.6

Switching on mains without selecting direction of travel/alarm volume

Switch on mains (O = OFF / I = ON).

The current software version is displayed for 10 seconds.

If no further buttons are pressed during this phase, after 10 sec-

onds the LED of the ON/OFF button lights up.

6.6.1 Stand-by mode

Press ‘ON/OFF’ button.

The fan goes on.

The LED (in the program button) of the program used last lights up
(here: ‘Prog 1’) and the programmed parameters are displayed.

Exception:
Prior to switching off the mains a program had not yet been fini-shed
or a power failure occurred while a program was in progress:

In those cases the instrument does not switch to the stand-by
mode but continues the previously interrupted program.