Instruction Manual V1.4 – 06/2009
4.1
Site requirements
The installation site has to meet the following
requirements:
•
Stable laboratory bench, exactly level, at
least 1.60 m wide and 60 cm deep for the ba-
sic instrument, plus an additional 0.30 m of
width for each of the optional stations.
•
Other than that, the same site requirements
as listed for the basic instrument apply (see
Chapter 3.1).
4.2
Unpacking the optional load /
unload station
The unpacking instructions for all Leica ins-
truments are located in a transparent pro-
tective envelope on the outside of the
instrument transportation crates.