4.
Installation of the Optional Load/Unload Station
•
Place grooved ramp no. 2 onto the rear pin
and use the hexagon key, size 1.5 to fasten
the two grub screws in the ramp (clockwise
rotation through about 180 °).
•
Place plain ramp no. 2 onto the front pin and
fasten the front ramp (clockwise rotation of
both grub screws through about 180 ° -
hexagon key, size 1.5).
Store ramps no. 1 for future use!
To operate the instrument with Medite
/ Hacker or Sakura slide racks, the
standard guide ramps no. 1 for Leica
slide racks must be exchanged for
guide ramps no. 2 (see ‘Standard deliv-
ery - optional unload station’, chapter
4.3.2).
•
Use the hexagon key, size 1.5 to unscrew
both ramps no. 1 loosening the two corre-
sponding grub screws in each ramp (coun-
terclockwise rotation through about 180 °).
Remove both ramps.
4.5
Changing the guide ramps for slide rack carriers
1
1
1
2
2
2