beautypg.com

3 cleaning and maintenance, Safety – Leica Biosystems ST4040 User Manual

Page 10

background image

10

Instruction Manual V1.4 – 06/2009

2.1.3 Cleaning and maintenance

Only technical service engineers authorized
by Leica may access the internal compo-
nents of the instrument for service and re-
pair.
Exception: Changing the activated carbon fil-
ter is the only maintenance task to be car-
ried out by the user.

Prior to cleaning the instrument, always
switch off mains and unplug the power
chord!

Dispose of used reagents according to the
laboratory regulations in force in your coun-
try!

Spilled solvents (reagents) have to be wiped
away immediately! - In case of long-term ex-
posure, the lid surfaces are only conditional-
ly resistant to solvents!

The painted surfaces and the control panel
are not resistant to xylene or acetone!

For cleaning the instrument, do not use any
one of the following: alcohol, detergents
containing alcohol (window cleaner!), abra-
sive cleaning powders, solvents containing
xylene or acetone!

To clean the lids, control panel and housing,
use mild household detergents; - see safety
instruction above for non-appropriate ingre-
dients!

When handling cleaning detergents, follow
the instructions of the manufacturer and
make sure all applicable laboratory regula-
tions are complied with!

When cleaning the instrument, no liquid may
enter in contact with any of the electrical
connections or get into the interior of the in-
strument!

Wash the tap water and reagent stations in
the dishwasher at a temperature of max. +65
°C. - Use a standard detergent for laboratory
dishwashers.
At any rate avoid washing the stations at
higher temperatures (e. g. in industrial dis-
washers which run at a temperature of
+85 °C), as the stations may become de-
formed!

2.

Safety