3 cleaning and maintenance, Safety – Leica Biosystems ST4040 User Manual
Page 10
10
Instruction Manual V1.4 – 06/2009
2.1.3 Cleaning and maintenance
•
Only technical service engineers authorized
by Leica may access the internal compo-
nents of the instrument for service and re-
pair.
Exception: Changing the activated carbon fil-
ter is the only maintenance task to be car-
ried out by the user.
•
Prior to cleaning the instrument, always
switch off mains and unplug the power
chord!
•
Dispose of used reagents according to the
laboratory regulations in force in your coun-
try!
•
Spilled solvents (reagents) have to be wiped
away immediately! - In case of long-term ex-
posure, the lid surfaces are only conditional-
ly resistant to solvents!
•
The painted surfaces and the control panel
are not resistant to xylene or acetone!
•
For cleaning the instrument, do not use any
one of the following: alcohol, detergents
containing alcohol (window cleaner!), abra-
sive cleaning powders, solvents containing
xylene or acetone!
•
To clean the lids, control panel and housing,
use mild household detergents; - see safety
instruction above for non-appropriate ingre-
dients!
•
When handling cleaning detergents, follow
the instructions of the manufacturer and
make sure all applicable laboratory regula-
tions are complied with!
•
When cleaning the instrument, no liquid may
enter in contact with any of the electrical
connections or get into the interior of the in-
strument!
•
Wash the tap water and reagent stations in
the dishwasher at a temperature of max. +65
°C. - Use a standard detergent for laboratory
dishwashers.
At any rate avoid washing the stations at
higher temperatures (e. g. in industrial dis-
washers which run at a temperature of
+85 °C), as the stations may become de-
formed!
2.
Safety