Instruction Manual V1.4 – 06/2009

4.4

Installation

4.4.1 Installing the optional load station

Remove the 4 hexagon socket screws from
the left lid (hexagon key, size 4).

Remove the lid. - Keep the screws (will be
needed later)!

Remove the 2 black screws from the left side
panel (hexagon key, size 3).

Remove the side panel from the side wall
and detach the ground wire from the side
wall.

Detach the ground wire and pull it out of the
instrument (will not be needed any longer.)

4.

Installation of the Optional Load/Unload Station

on the left side of the basic instrument

Press the Start/Stop button, to move the
conveyor frame to the upper position. As
soon as the conveyor frame has reached the
upper end position, switch the instrument off
(ON/OFF button).