Leica ST4040 – Linear Stainer
3.6.2 Installing the drain hose
•
Connect the drain hose.
Upon installing the drain hose make
sure there is a gradient from the drain
outlet to the waste pipe.
3.6.3 Installing the exhaust air hose
•
Connect the exhaust air hose (optional!).
The instrument may be operated ei-
ther with the exhaust air hose or
with the activated carbon filter.
3.5.2 Connecting the power chord to the
•
Remove the adhesive tape from the mains
inlet.
•
Select the appropriate power chord (the in-
strument is delivered with several, country-
specific power chords) and connect it to the
mains inlet at the rear of the instrument.
•
Do not yet plug the power chord into the wall
outlet.
•
The unload sensor port remains empty; un-
less the instrument is operated with optional
unload station (see Chapter 4).
3.
Installation
3.6
Installing the accessories
3.6.1 Installing the water inlet hose
•
Install the inlet hose, which supplies water
to the tap water stations.
instrument