beautypg.com

3 lubricating the instrument, Cleaning and maintenance – Leica Biosystems RM2245 User Manual

Page 72

background image

70

Instructions for Use V 1.9 RevC - 06/2013

8. Cleaning and Maintenance

97

88

11

10

Knife holder E:
• Clamping lever (11) for the lateral displace-

ment.

• T-piece (88) and guide (97) of the knife holder

for the lateral displacement.

• Clamping lever (10) for the blade.

Knife holders N and NZ:
• Sliding surfaces of the finger guard (8).
• Knurled nuts (73) for measuring height adjust-

ment.

95

96

55

8.2.3 Lubricating the instrument

Once a month, lubricate the following parts with

the included drive parts oil No. 405 (1 - 2 drops

are enough).

Instrument and specimen holder:
• The clamping key (95) of the quick clamping

system.

• The T-piece (55) on the microtome base plate.
• Guide rails (96) for the knife holder base on the

microtome base plate.

• The T-piece on the knife holder base (56).

56

73

8

Knife holder N

Knife holder NZ

Fig. 74