8. Cleaning and Maintenance

Knife holder E

10

11

9

86

83

83

86

87

11

10

Take the knife holder apart for cleaning. To do so,

proceed as follows:

• Fold the knife guard (9) and ejector (84) downward.

• Rotate the clamping lever (11) of the lateral dis-

placement forwards and pull it out sideways.

• Push the knife holder base plate (86) with the pres-

sure plate (83) until it can be removed from the

segment arch (87).

• Rotate the blade clamping lever (10) downward and

pull it out sideways.

• Remove pressure plate (83).

• Clean all parts of the knife holder.

84

There is a burn hazard when removing the

parts from the drying chamber (65 °C). Wear-

ing safety gloves is recommended!

• Dry knife holder and reassemble.
• After cleaning the moving parts of the knife holder,

apply a thin coat of drive part oil to them (see also

Chap. 8.2.3

).

• When installing the pressure plate (83) ensure that

the upper edge is parallel to and level with the rear

edge of the knife holder base (86) (

refer also to

Chap. 6.3.2, Fig. 43

).

If several knife holders are cleaned at the

same time, the parts must NOT be mixed up!

Failure to adhere to this may result in sec-
tioning problems!

For cleaning and removal of paraffin, do not

use xylene or cleaning fluids containing

alcohol (e.g. glass cleaner).

• Lay the removed parts on an absorbent cloth into

the drying chamber (up to a max. of 65 °C) and let

the paraffin contamination run off.

Fig. 70

Fig. 71