7. Troubleshooting

7.1.1 Error reports

Indication E1
If this message follows the software version indication after switching on

the instrument, a loss of all parameters has occurred.
• Press any key.
All parameters are then set back to the default values.
• Now all settings must be checked to see if the previously entered values

are still correct.

The values may have to be reentered.
You can continue with your normal working routine.

If E-1 appears repeatedly when switching the instrument on, the lithium

battery may need replacement (service life approx. 7 years).
• Proceed as described above.
• Work normally after checking the settings.
• Have the battery replaced by Leica service as soon as possible.

LEDs (40), (41) in the coarse feed buttons are lit at the same time.
This indicates a defect in the detection of the STOP and HOME positions of

the specimen head.
The problem can only be resolved by customer service.

7.1 Instrument malfunctions

40

41

Certain instrument malfunctions appear on the 3-digit LED display or are indicated

through illuminated LED's on the control panel.