Troubleshooting, 1 instrument malfunctions, 1 error reports – Leica Biosystems RM2245 User Manual
Page 64: Instrument malfunctions
62
Instructions for Use V 1.9 RevC - 06/2013
7. Troubleshooting
7.1.1 Error reports
Indication E1
If this message follows the software version indication after switching on
the instrument, a loss of all parameters has occurred.
• Press any key.
All parameters are then set back to the default values.
• Now all settings must be checked to see if the previously entered values
are still correct.
The values may have to be reentered.
You can continue with your normal working routine.
If E-1 appears repeatedly when switching the instrument on, the lithium
battery may need replacement (service life approx. 7 years).
• Proceed as described above.
• Work normally after checking the settings.
• Have the battery replaced by Leica service as soon as possible.
LEDs (40), (41) in the coarse feed buttons are lit at the same time.
This indicates a defect in the detection of the STOP and HOME positions of
the specimen head.
The problem can only be resolved by customer service.
7.1 Instrument malfunctions
40
41
Certain instrument malfunctions appear on the 3-digit LED display or are indicated
through illuminated LED's on the control panel.