3 technical data, Technical data, Instrument components and specifications – Leica Biosystems RM2245 User Manual
Page 15
13
Leica RM2245
3.3 Technical Data
General
Approvals:
The instrument-specific marks are located next to the
identification label.
Nominal supply voltages:
100/120/230/240 V AC ±10 %
Nominal frequency:
50/60 Hz
Maximum power consumption:
70 VA
Protection class
: I
Power fuses:
2 x T 1.0 A UL listed
Pollution degree
: 2
Overvoltage category
: II
Maximum heat emission:
70 J/s
Operating temperature range:
+10 °C to +35 °C
Temperature range during storage:
+5 °C to +55 °C
Relative humidity:
max. 80 % non-condensing
Storage humidity:
< 80 %
according to IEC-1010, UL 3101, EN 61010
Dimensions and weight
Basic instrument
Width (including handwheel):
413 mm
Width (excluding handwheel):
300 mm
Depth (including waste tray):
618 mm
Height (total):
305 mm (with tray on the hood)
Working height (knife blade):
100 mm (measured from the base plate)
Working height (knife blade):
168 mm (measured from the table)
Weight (without accessories):
approx. 37 kg
Control panel
Width:
94 mm
Depth:
164 mm
Height:
50 mm
Height (in inclined position):
81 mm
Weight (net):
approx. 0.450 kg
3. Instrument Components and Specifications