31

Leica RM2245

5. Operation

There are two versions of the specimen holder

fixture, one with and one without specimen ori-

entation, which are interchangeable.

The specimen orientation allows for simple posi-

tion correction of the specimen surface when the

specimen is clamped into place.

You can use the quick clamping system (64) to

hold all available accessory specimen clamps

(for more information, see

Chapter 6

"Optional

accessories").

To do so, proceed as follows:
• Move the specimen head (60) to the upper end

position by turning the handwheel (12) and

engage the handwheel lock.

• To release the clamping system, turn the

screw (61) of the quick clamping system (64)

counterclockwise using an Allen key No. 4 (71).

• Push the guide (63) of the universal cassette

clamp (62) from the left into the quick clamping

system (64) as far as it will go.

• To clamp the cassette clamp turn the screw

(61) clockwise as far as it will go.

5.3 Inserting the universal cassette clamp

71

63

62

60

12

61

64

60

Since all stage clamps available as accessories are equipped with the same kind of guide

on the back, they are inserted in the same way described here using the example of the

cassette clamp.

Fig. 24