beautypg.com

3 knife holder base and knife holder, Knife holder base and knife holder, Optional accessories – Leica Biosystems RM2245 User Manual

Page 48

background image

46

Instructions for Use V 1.9 RevC - 06/2013

6. Optional Accessories

6.3 Knife holder base and knife holder

Repositioning the knife holder base
The one-piece knife holder base (rigid) (51) can be

moved forwards and backwards on the microtome

base plate.
This vertical displacement allows bringing the

knife holder into the optimal cutting position in

relation to the specimen.

• To release, rotate the clamping lever (50) on

the right side of the microtome base plate

counterclockwise.

• Reposition the knife holder together with the

knife holder base forward or backward as

appropriate.

• Secure the clamping mechanism by rotating

the lever (50) clockwise.

The plastic handles of all clamping levers on the

instrument and knife holders can be turned to the

position that is most convenient for each user.


Pull the grip (94) out of the lever, hold it in this

position, and rotate it to the desired position. It

will then lock automatically when released.

6.3.1 Knife holder base, without lateral

displacement feature

51

50

94

Fig. 40

Fig. 39