1 replacing fuses, 2 maintenance, Cleaning and maintenance – Leica Biosystems RM2245 User Manual

Page 70

background image

68

Instructions for Use V 1.9 RevC - 06/2013

Universal cassette clamp

• Detach cassette clamp (13) for a thorough cleaning, removing all paraffin

residues.

• For cleaning, do not use xylene. Use xylene substitutes or paraffin remov-

ers such as "Para Gard."

• The cassette clamp (13) can also be placed in an oven heated to a maxi-

mum of 65 °C, until the liquid wax escapes.

• Remove paraffin residues with a dry cloth.
• After such a cleaning procedure in an oven, always be sure to lubricate

the spiral springs (91) of the tension lever (60) and move it back and forth

several times.

8. Cleaning and Maintenance

• Insert a small screwdriver into the cutout (25)

and carefully pry out the insert.

• Remove the voltage selector housing (21)

together with the fuses (23).

• Remove the fuses. While doing so, leave the

voltage selector switch block (24) in the hous-

ing.

• Replace defective fuses and reinstall the volt-

age selector into the instrument.

• Check that the correct voltage value is dis-

played in the viewing window (22).

8.2.1 Replacing fuses

8.2 Maintenance

22

25

24

21

23

21

60

13

91

Fig. 72

Fig. 73

Switch the unit off and unplug it. Use

only the supplied replacement fuses.

Both fuses must have the same rating

(check the imprint!).