Instrument setup – Leica Biosystems RM2245 User Manual
Page 21
![background image](/manuals/565481/21/background.png)
19
Leica RM2245
4. Instrument Setup
22
25
24
21
23
The voltage selector is located above the main
power switch, on the left-rear side of the instru-
ment (
Fig.
7). The voltage setting is displayed in
the viewing window (22).
• Insert a small screwdriver into the cutout (25)
and carefully pry out the insert.
• Remove the voltage selector housing (21) to-
gether with the fuses (23). Remove the voltage
selector block (24) (white) and insert it again
such that the correct local voltage is displayed
in the viewing window (22).
• Reinsert the voltage selector housing with the
block and fuses and push it in until it engages
(audible click).
21
26
27
Checking the voltage
(continued)
Connecting the control panel
The connecting cable (15) of the control panel
is fixed to the microtome. It may not be discon-
nected.
• Insert the plug (14) of the connecting cable
(15) into the connecting jack (16) on the rear
of the control panel.
• To fasten the plug, tighten the two screws (17).
15
16
14
17
Fig. 12
Fig. 13