beautypg.com

2 trimming, Trimming, Operation – Leica Biosystems SM2500 User Manual

Page 61

background image

61

Leica SM2500

5.6.2 Trimming

Trimming can be done in modes of operation AUTO or PROG.

Trimming in AUTO mode of operation:

Preparatory steps:

Activate AUTO mode of operation (see chapter 5.4.3).

Select the desired section thickness (see chapter 5.4.5).

Activate retraction / select retraction setting as needed (chapter 5.4.5).

Select sectioning / milling and return stroke speed (see chapter 5.4.7).

Select a sectioning mode (see chapter 5.4.6) (for trimming usually
CONT).

Turn the knife guard away from the knife edge.

Start the trimming procedure:

Press RUN STOP + RUN ENABLE (see chapter 5.4.10) or press the foot
switch (see chapter 5.4.1).

Trim the specimen down to the desired level.

Press RUN STOP or RUN ENABLE or once again press the foot switch to
stop trimming.

Trimming in PROG mode of operation

Preparatory steps:

Activate PROG mode of operation.

The difference between PROG and AUTO is that in PROG none of
the three sectioning modes (area 2.3, chapter 5.4.6) is selected, but
a programmed number of sections is carried out.

Select section thickness, retraction (if needed) and speed as described
for AUTO mode of operation.

Select the desired number of sections ‘n’ you wish to be carried out
(area 2.2 chapter 5.4.5).

Start the trimming procedure:

Press RUN STOP + RUN ENABLE or press the foot switch:

the instrument carries out the preselected number of sections and then
stops in the selected stop or photo position.

5.

Operation