69

Leica SM2500

Replacing the vacuum cleaner or lamp fuse

Insert a screwdriver into slot (1); push fuse
holder (2) slightly inwards turning at the same
time by 1/4 turn to the left.

The fuse holder is released and can be re-
moved.

Replacing the line fuses

To replace the line fuses, proceed exactly as
per chapter 4.5.1 ‘Verifying and adjusting the
voltage selector’.

Do not, however, change the orientation of the
voltage selector, but simply replace the defec-
tive fuses by new ones of correct specifica-
tion.

Remove the defective fuse (3) from the fuse
holder (2) and replace by a replacement fuse
of correct specification.

Reinsert the fuse holder / fuse into the instru-
ment. Use the screwdriver to push slightly in-
wards, turning at the same time by 1/4 turn to
the right.

7.

Cleaning, disinfection and maintenance

Caution!
After replacing the line fuses and rein-
serting the voltage selector housing
into the instrument, double-check
whether the correct voltage selector
setting is visible.

1

2

Fig. 76

2

3

Fig. 77