beautypg.com

4 description of each individual control element, 1 foot switch, Description of each individual control element – Leica Biosystems SM2500 User Manual

Page 24: Foot switch, Operation

background image

24

Instruction manual V 1.3 – 10/2007

5.

Operation

5.4

Description of each individual control element

5.4.1 Foot switch

The foot switch has two different functions:

1 Start / stop sectioning / milling.

The foot switch can be used as an alternative to the RUN STOP + RUN
ENABLE buttons (--> see area 2.6, chapter 5.4.10) so start / stop section-
ing / milling.

- To start, lightly press the foot switch.

- To stop, once again lightly press the foot switch.

2 Emergency stop.

In addition to the emergency stop button (see area 1, chapter 5.4.2), the
emergency stop can be activated via the foot switch:

-

Press foot switch once forcefully: emergency stop function is activated.
POWER (green) and E-STOP (red) on the control emit light.

As soon as the foot switch is released the emergency lock is auto-
matically unlocked and the instrument can be restarted.
E-STOP (red) on the control panel goes off.

Fig. 14