beautypg.com

Operation – Leica Biosystems SM2500 User Manual

Page 41

background image

41

Leica SM2500

Clamping cylinder-shaped specimen blocks

Remove screws (6) holding the fixed clamp-
ing jaw (1) (Allen key, size 5).

Turn the jaw and re-insert with the V-notch
facing inwards, towards the specimen.

Re-insert and tighten screws (6) of clamping
jaw (1).

Insert the movable jaw, (2) in a position that is
appropriate for the specimen size to be
clamped and with the V-notch facing inwards
(towards the specimen).

Tighten the movable jaw via clamping screw
(3), using an Allen key size 8 or a single-head
wrench, size 13.

Important - for square as well as for
cylindrical specimen blocks:
When the specimen block is tightly
clamped, the front edge of block (4)
must not have moved beyond the two
lateral markings (5).

Clamping specimen stages

The fixed clamping jaw (1) remains in standard
position (V-notch facing outwards, i.e. away
from the specimen stage).

All specimen stage sizes (plastic and metal spec-
imen stages) have a clamping block of identical
size and are clamped as follows:

Insert the movable jaw (2) with its front edge
(i.e. the edge facing the specimen) flush with
marking (7).

Insert the specimen stage (e.g. 8, 9).

Tighten the movable jaw via clamping screw
(3), using an Allen key size 8 or a single-head
wrench, size 13.

5.

Operation

2

1

3

5

4

5

6

7

7

Fig. 39

8

Fig. 40

9

Fig. 41