7 area 2.4 - sledge speed, Area 2.4 - sledge speed, Operation – Leica Biosystems SM2500 User Manual
Page 30
30
Instruction manual V 1.3 – 10/2007
5.4.7 Area 2.4 - Sledge speed
19 SPEED (button)
To set the cutting or return speed, press and hold the SPEED button until
the corresponding LED blinks. Enter speed using the rotary button (9).
20 CUT (LED)
Press and hold SPEED until the LED CUT starts blinking. Then enter the
desired sectioning / milling speed of the sledge via selector switch (9)
(selectable speeds: 0.5 - 100 mm/sec, in 0.1-mm steps).
21 RETURN (LED)
Press and hold SPEED until the LED RETURN starts blinking. Then enter
the desired return stroke speed of the sledge via selector switch (9) (se-
lectable speeds: 0.5 - 100 mm/sec, in 0.1-mm steps).
22 Speed indication
Indicates sledge speed during sectioning / milling stroke and return
stroke.
CUT and RETURN can be changed while sectioning / milling is in
progress.
All settings entered are automatically stored upon entry.
No additional confirmation is required.
The LEDs stop blinking automatically after a few seconds.
20
22
19
21
5.
Operation
Fig. 20