8 Supplement
Resistance transmitter
Voltage
Measuring circuit monitoring
8.1.2
Output (radio transmission)
Designation
Measuring range
Measuring accuracy
a
Resistance transmitter
>
50
Ω
to
≤
400
Ω
>
400
Ω
to
≤
4000
Ω
>
4000
Ω
to
≤
10000
Ω
±
400 m
Ω
±
4
Ω
±
10
Ω
Connection type
3-wire connection
Requirement
R
e
+ R
a
+ R
s
≤
10000
Ω
and R
e
+ R
a
≤
1/3 R
s
a
All accuracies given in % refer to the maximum measuring range
Designation
Measuring range
Measuring accuracy
a
Voltage
0 to 50 mV
±
0.1 %
a
All accuracies given in % refer to the maximum measuring range
Measuring probe
Detection of
overrange / under-
range
Detection of
probe /
cable short-circuit
Detection of
probe /
cable break
Thermocouple
Yes / Yes
No
Yes
RTD temperature probe
Yes / Yes
Yes
Yes
Resistance / potentiometer
Yes / Yes
Yes
Yes
Resistance transmitter
No / No
No
No
Voltage
Yes / Yes
No
Yes
Probe ID
Max. 5-digit ID, factory-set, can be configured to customer-request
Transmission interval
Adjustable from 1 to 3600 s (factory setting: 15 s)
Transmission frequency
868.4 MHz (Europe)
Transmission power
+10 dBm
Open air range
Max. 300 m when using the antenna wall-holder of the receiver and
the 3 m antenna line. When installing the antenna directly on the
receiver, a reduced range of approx. 40 % must be taken into
account.
Output signal
Thermocouple
Voltage (mV)
RTD temperature probe
Resistance (
Ω
)
Resistance / potentiometer
Percent (%) or resistance (
Ω
)
Resistance transmitter
Percent (%)
Voltage
Voltage (mV)