beautypg.com

8 supplement, 1 technical data, 1 analog input – JUMO 707060 Wtrans B - Programmable Head Transmitter with Radio Transmission Operating Manual User Manual

Page 37: Supplement, Technical data

background image

37

8 Supplement

8.1

Technical data

8.1.1

Analog input

RTD temperature probe

Designation

Standard

Measuring range

Measuring accuracy

Pt100

DIN EN 60751

-100 to +200 °C

± 0.1 K

(TK value = 3.85 × 10

-3

1/K)

-200 to +600 °C

±

0.2 K

Pt500

DIN EN 60751

-100 to +200 °C

± 0.1 K

(TK value = 3.85 × 10

-3

1/K)

-200 to +600 °C

±

0.2 K

Pt1000

DIN EN 60751

-100 to +200 °C

± 0.1 K

(TK value = 3.85 × 10

-3

1/K)

-200 to +600 °C

±

0.2 K

Ni 100

DIN 43760

-60 to +250 °C

±

0.2 K

(TK value = 6.18 × 10

-3

1/K)

Ni 500

DIN 43760

-60 to +150 °C

±

0.2 K

(TK value = 6.18 × 10

-3

1/K)

Ni 1000

DIN 43760

-60 to +150 °C

±

0.2 K

(TK value = 6.18 × 10

-3

1/K)

Pt100

JIS 1604

-100 to +200 °C

± 0.1 K

(TK value = 3.917 × 10

-3

1/K)

-200 to +600 °C

±

0.2 K

Pt50

ST RGW 1057 19
85

-200 to +600 °C

±

0.2 K

(TK value = 3.91 × 10

-3

1/K)

Pt100

GOST 6651-
94 A.1

-100 to +200 °C

± 0.1 K

(TK value = 3.91 × 10

-3

1/K)

-200 to +600 °C

±

0.2 K

Cu50

GOST 6651-
94 A.4

-50 to +200 °C

±

0.2 K

(TK value = 4.26 × 10

-3

1/K)

Cu100

GOST 6651-
94 A.4

-50 to +200 °C

±

0.2 K

(TK value = 4.26 × 10

-3

1/K)

Connection type

2-wire or 3-wire circuit

Sensor current

< 0.5 mA

Lead compensation

Not required for 3-wire circuit (admissible max. 11

Ω

per line).

Max. configurable lead-wire resistance for 2-wire circuit: 22

Ω

.