beautypg.com

2 fine adjustment, 6 configuration – JUMO 705020 mTRON T - Analog Input Module, 4-Ch. Operating Manual User Manual

Page 41

background image

41

6 Configuration

6.2.2

Fine adjustment

You can use customer-specific fine adjustment to correct the measured values of the analog
input. In contrast to offsetting, which is used to specify a constant correction value for the entire
characteristic line, fine adjustment can also be used to change the gradient of the characteristic
line.

Setup dialog

Parameter

Example

The temperature inside a furnace is measured with an RTD temperature probe and displayed.
Due to the temperature drift of the sensor, the true temperature (reference measurement) de-
viates from the displayed value. The amount of deviation is different at the upper and lower
measuring points, meaning that measured value offset is not suitable.
Actual start value: 15 °C (displayed value)
Target start value: 20 °C (reference measurement)
Actual end value: 70°C (displayed value)
Target end value: 80 °C (reference measurement)

Parameter

Selection/settings

Description

Actual start value -99999 to 0 to +99999

Lower display value

Target start value -99999 to 0 to +99999

Lower reference value

Actual end value

-99999 to 1 to +99999

Upper display value

Target end value

-99999 to 1 to +99999

Upper reference value

20°C / 80°C

20°C / 80°C

15°C / 70°C

(4)

(3)

(1)

(2)

(1)

Display value

(2)

Reference value

(3)

Furnace

(4)

Sensor in RTD temperature probe