JULABO LC 6 Programmable Controller User Manual
Page 48

Remote control
48
Command
Parameter
Response of circulator
out_mode_02
2
Continual identification of controlled system whenever a new
setpoint is to be reached.
out_mode_04
0
Temperature control with Pt100 sensor „INT“.
out_mode_04
1
Temperature control with Pt100 sensor „EXT“.
out_mode_05
0
Stop the programmable controller = r OFF.
out_mode_05
1
Start the programmable controller.
out_sp_00
xxx.x
Set working temperature "T1".
out_sp_01
xxx.x
Set working temperature "T2".
out_sp_03
xxx.x
Set high temperature warning limit
.
out_sp_04
xxx.x
Set low temperature warning limit
.
out_hil_00
xxx
Set the desired maximum cooling power (0 % to 100 %).
This adjustment is required only for proportionally controlled
refrigerated circulators.
out_hil_01
xxx
Set the desired maximum heating power (10 % to 100 %).
out_par_04
xxx
Band limiting during external control.
Setting the maximum difference between the temperatures
in the internal bath and external system.
out_par_05
xxx
Ratio for max. cooling power versus max. heating power
(0...0.99).
out_par_06
xxx
Xp control parameter of the internal controller.
out_par_07
xxx
Tn control parameter of the internal controller.
out_par_08
xxx
Tv control parameter of the internal controller.
out_par_09
xxx
Xp control parameter of the cascade controller.
out_par_10
xxx
Proportional portion of the cascade controller.
out_par_11
xxx
Tn control parameter of the cascade controller.
out_par_12
xxx
Tv control parameter of the cascade controller.