Menu functions

38

8.6. Limits

When operating the programmable controller under external
control, band limiting is active. The preset value determines the
maximum temperature difference between the internal bath
and the external load. This adjustment possibility prevents
sensitive equipment and temperature devices from damage.

Heating and cooling power of the programmable controller are
adjustable.
100 % corresponds to the values in the technical specifications
of the equipment.

 Select the submenu „Limits“ with enter

.

 Select the desired option with the up/down cursor keys

.

A flashing digit indicates that a value needs to be entered.
Band

0 to 200 °C

HeatingMax

0 to 100 % in steps of 1 %

CoolingMax

0 to 100 % in steps of 1 %

 To set the newly entered value press enter

.

 To return to the previous menu level press escape

.

8.7. Interface

The interface parameters are set by selecting the submenu
„Interface“ on the programmable controller. Normally, this is a
one-time-only adjustment.

 Press enter

to select the submenu „Interface“.

 Select the desired option with the up/down cursor keys

.

Enter the desired value for the flashing digit.

Type

RS232 / RS485

Baudrate 2400/4800/9600
Parity

none/even/odd

Handshake

software handshake/hardware handshake


Address

0 to 127

 Press the P-key

to select the desired parameter and

set with enter

.

 Return to the previous menu level with escape

.