Menu functions
38
8.6. Limits
When operating the programmable controller under external
control, band limiting is active. The preset value determines the
maximum temperature difference between the internal bath
and the external load. This adjustment possibility prevents
sensitive equipment and temperature devices from damage.
Heating and cooling power of the programmable controller are
adjustable.
100 % corresponds to the values in the technical specifications
of the equipment.
Select the submenu „Limits“ with enter
.
Select the desired option with the up/down cursor keys
.
A flashing digit indicates that a value needs to be entered.
Band
0 to 200 °C
HeatingMax
0 to 100 % in steps of 1 %
CoolingMax
0 to 100 % in steps of 1 %
To set the newly entered value press enter
.
To return to the previous menu level press escape
.
8.7. Interface
The interface parameters are set by selecting the submenu
„Interface“ on the programmable controller. Normally, this is a
one-time-only adjustment.
Press enter
to select the submenu „Interface“.
Select the desired option with the up/down cursor keys
.
Enter the desired value for the flashing digit.
Type
RS232 / RS485
Baudrate 2400/4800/9600
Parity
none/even/odd
Handshake
software handshake/hardware handshake
Address
0 to 127
Press the P-key
to select the desired parameter and
set with enter
.
Return to the previous menu level with escape
.