8.3.

Start of a profile









1.


2.

The start menu of the programmable programmer allows
calling up and defined starting of one of six previously stored
temperature profiles.

There are two possibilities for manually starting a
program:

1. Starting a program from the OFF status:

The programmer switches back to the OFF status at the
end of the program.

2. Starting a program from the operating status.

The programmer is started with the Start key

, and the

bath is heated to the desired temperature, for example
100 °C. At the end of the program, the programmer
switches to the operating status and holds the bath
temperature stable at 100.00 °C.



 Press enter

to select the submenu „Profile Start“.

 Use the up/down cursor keys

to select the

desired option.

A flashing segment indicates that a number needs to be
entered.
Start Profile

0 to 5

at Step

0 to 60

Loops

1 to 99

Enter the desired number and set each entry with enter

.


Start

no / yes

 (m a n u a l s t a r t )

or

time

 (via integrated timer)

A flashing line indicates that a parameter needs to be
entered.

Press the P-key

to select the respective parameter

and press enter

.