8.8. Sensors



ATC - Absolute Temperature Calibration

 Select the submenu „Temp.Sensor“ with enter

.

 Select the desired option with the up/down cursor keys

.

A flashing digit indicates that a value needs to

be entered i.e. set.
ATC Int.: used during the configuration of a Pt100 sensor
attached on the socket labeled “INT”.

ATC Ext.: used during the configuration of a Pt100 sensor

attached on the socket labeled “EXT”.

Setpoint range for internal or external calibration: ±9.99 °C.

 Enter the desired correction value and set this value by

pressing ENTER

.

Thermometer (T

M

)

Programmable controller (T

L

)

Example:

1. Immerse temperature sensors individually or together in a

calibration bath at 50°C, for example, and allow the reading
to stabilize.

2. Both temperature values are displayed simultaneously on

the controller readout:
on the MULTI-DISPLAY:

49.20°C for temperature sensor attached to the socket

labeled

“INT”.

on DIALOG-DISPLAY:

51.22°C for temperature sensor attached to the socket

labeled

“EXT”.

3. Calculate the temperature difference between the

programmable controller (T

L

) and the thermometer (T

M

)

(change in T = T

M

- T

L

) and enter the correction value as the

ATC parameter.

4. For ATC Int, enter the correction value (for example

0.80 °C) with the keypad. Confirm entry with the Enter key

.

5.

6. For ATC Ext, enter the correction value,

(for example -1.22 °C) followed by the Enter key

.

7. Return to standard display with the Escape key

.