Operator responsibility – Safety recommendations

10


Safety installations according to IEC 61010-2-010:

Excess temperature protection

adjustable from 0 °C ... 420 °C

Supplementary safety installations

High temperature warning function

optical + audible (in intervals)

Low temperature warning function

optical + audible (in intervals)

Supervision of the working sensor

plausibility control

Alarm indication

optical + audible




Environmental conditions according to EN 61 010, part 1:

Use only indoor.

Altitude up to 2000 m - normal zero.

Ambient temperature: +5 ... +40 °C (for storage and transportation)

Air humidity:

Max. rel. humidity 80 % for temperatures up to +31 °C,

linear decrease down to 50 % relative humidity at a temperature of +40 °C

Max. mains fluctuation of ±10 % are permissible.


Protection class according to EN 60 529

IP31

The unit corresponds to Class I

Overvoltage category II

Pollution degree 2

Caution:

The unit is not for use in explosive environment.


Standards for interference resistance according to EN 61326-1
This unit is an ISM device classified in Group 1 (using high frequency for internal purposes) Class