JULABO LC 6 Programmable Controller User Manual

Page 47

background image

47

Command

Parameter

Response of programmable controller

in_mode_04

none

Internal/external temperature control:

0 = Temperature control with Pt100 sensor „INT“.

1 = Temperature control with Pt100 sensor „EXT“.

in_mode_05

none

Circulator in Stop/Start condition:

0 = Stop

1 = Start

in_par_01

none

Time constant of the external bath.

in_par_02

none Internal

slope.

in_par_03

none

Time constant of the internal bath.

in_par_04

none

Band limiting (max. difference between the temperatures in
the internal bath and external system).

in_par_05

none

Ratio for max. cooling power versus max. heating power.

in_par_06

none

Xp control parameter of the internal controller.

in_par_07

none

Tn control parameter of the internal controller.

in_par_08

none

Tv control parameter of the internal controller.

in_par_09

none

Xp control parameter of the cascade controller.

in_par_10

none

Proportional portion of the cascade controller.

in_par_11

none

Tn control parameter of the cascade controller.

in_par_12

none

Tv control parameter of the cascade controller.





out commands: Setting parameters or temperature values.

Command

Parameter

Response of circulator

out_mode_01

0

Use working temperature "T1"

out_mode_01

1

Use working temperature "T2"

out_mode_02

0

No identification.
Temperature control by using the stored parameters.

out_mode_02

1

Single identification of controlled system after the next start.