beautypg.com

JULABO LC 6 Programmable Controller User Manual

Page 27

background image

27


Autostart

Note:
The programmable controller has been configured and
supplied by JULABO according to N.A.M.U.R.
recommendations. This means for the start mode, that the unit
must enter a safe operating state after a power failure (non-
automatic start mode). This safe operating state is indicated by
„OFF“ or „rOFF“, resp. on the MULTI-DISPLAY (LED). A
complete shutdown of the main functional elements such as
heater and circulation pump is effected simultaneously.

Should such a safety standard not be required, the
AUTOSTART function (automatic start mode) may be
activated, thus allowing the start of the programmable
controller directly by pressing the mains power switch or using
a timer.

Possible parameters:

on - AUTOSTART on

off - AUTOSTART off

Warning

:

For supervised or unsupervised operation with the AUTOSTART function, avoid
any hazardous situation to persons or property.
The programmable controller does no longer conform to N.A.M.U.R.
recommendations.

Take care you fully observe the safety and warning functions of the
programmable controller.

STAND-BY

Stand-by input

External stand-by for emergency switch-off
(connector - see page 6)

Possible parameters:

no - stand-by input is ignored

yes

- stand-by input is active

Language

There are two options for the language of the DIALOG-
DISPLAY (LCD): German or English.

Possible parameters:

German (deutsch)

English (engl.)