JULABO LC 6 Programmable Controller User Manual

Page 28

background image

Menu functions

28

8.2. Control

parameters




Xp =

When performing an identification for the controlled system
(temperature applications system) (see page 25), the control
parameters Xp, Tn, and Tv will be automatically determined
and stored.

Each parameter may be manually set via the keypad if
necessary, to allow optimum control performance.

 Press enter

to select the submenu „control

parameters“.

 Use the up/down cursor keys

to select the

desired option. A flashing segment indicates that a new
value needs to be entered.

 Use the numeral keypad to set the value and then set with

enter

(example: Xp = 2.5 °C).

 Press escape

to return to the previous menu level.

Optimization instructions for the PID control parameters:

The heat-up curve reveals inappropriate control settings.
(example: working temperature T1)

optimum setting

Inappropriate settings may produce the following
heat-up curves:



Xp too low



Tv/Tn too low



Xp too high
or
Tv too high



Tv/Tn too high
or
Xp too high