8.1. Configuration
By means of the configuration functions, operation of the
programmable controller can be optimized for the current
application.
Press enter
to select the configuration submenu.
Use the up/down cursor keys
to select the
desired option. A flashing line indicates that a value needs
to be entered.
Press the P-key
to select the parameter and press
enter
.
Press escape
to return the previous menu level.
Identification
When performing an identification for the controlled system
(temperature application system), the control parameters Xp,
Tn and Tv will be automatically determined and stored.
Possible parameters:
off - no identification.
The control parameters ascertained during the last
identification are used for control purposes.
once - single identification
The programmable controller performs a single
identification of the controlled system after start.
After the identification process the parameter is
automatically set to “off“.
always - continual identification
The programmable controller performs an identification of
the controlled system whenever a new setpoint is to be
reached.
NOTE: Use this setting only when the temperature
application system changes permanently.