Operating procedures – JULABO LC 6 Programmable Controller User Manual
Page 15

15
5. Operating
procedures
5.1. Installation
The unit should be set up at a dry location.
Place the unit in an upright position and do not obstruct the
ventilation.
A wall distance of at least 10 cm must be maintained for
ventilation, allowing internal heat
to be conducted away from the unit.
If one or more temperature controllers are set up in a
cabinet for example, take care of good ventilation (waste
heat per unit = approx. 60 Watts).
The ambient temperature must not exceed 35 °C. Ambient
temperatures above 35 °C result in a failure of the unit.
Do not set up the unit in the immediate vicinity of heat
sources and do not expose to sun light.
5.2.
Connecting a heating device
Caution:
Set up the heating device according to the instructions or securely fix the unit in the
bath tank using appropriate means. Danger of burning and fire!
Connect the power plug to the grounded mains socket (36) on
the rear of the programmable controller.
Caution:
Max. resistive load 3000 W at 230 V. Max. current 16 A.
Max. resistive load 1400 W at 115 V. Max. current 12 A.
+
- AWC100 Compact Recirculating Cooler (15 pages)
- CF30 Cryo-Compact Circulators (35 pages)
- CF31 Cryo-Compact Circulators (76 pages)
- ED Heating Immersion Circulator (23 pages)
- ED-5A/B Bath Circulators (24 pages)
- ED-5A Open Bath Circulators (24 pages)
- EH Heating Immersion Circulator (27 pages)
- EH-5 Open Bath Circulator (28 pages)
- F250 Recirculating Cooler (35 pages)
- F12-ED Refrigerated / Heating Circulator (26 pages)
- F12-EH Refrigerated and Heating Circulators (32 pages)
- F12-MA Refrigerated and Heating Circulators (65 pages)
- F38-ME Beer Forcing Test Refrigerated/Heating Circulating Bath (69 pages)
- F38-EH Refrigerated and Heating Circulators (30 pages)
- F25-ME Refrigerated and Heating Circulators (66 pages)
- F25-HE Refrigerated and Heating Circulators (86 pages)
- F34-ED Refrigerated / Heating Circulator (25 pages)
- F25-HL Refrigerated Circulators (94 pages)
- FL7006 Recirculating Coolers (34 pages)
- FL2503 Recirculating Coolers (36 pages)
- FL20006 Recirculating Coolers (32 pages)
- FL1201 Recirculating Coolers (33 pages)
- FL300 Recirculating Coolers (48 pages)
- FK30-SL Calibration Baths (88 pages)
- FT200 Immersion Coolers (20 pages)
- FC600_600S Recirculating coolers (37 pages)
- FC1200S Recirculating cooler (45 pages)
- F81-HL Ultra-Low Refrigerated Circulator (93 pages)
- F70-ME Ultra-Low Refrigerated Circulators (64 pages)
- FP52-SL Ultra-Low Refrigerated Circulators (103 pages)
- FP51-SL Ultra-Low Refrigerated Circulators (90 pages)
- FP45-HE Refrigerated and Heating Circulators (79 pages)
- HL-4 Heating Circulators (89 pages)
- HE-4 Heating Circulators (78 pages)
- F81-HL Ultra-Low Refrigerated Circulators (93 pages)
- FT402 Immersion Coolers (24 pages)
- ME-4 Heating Circulators (62 pages)
- ME-31A Visco-Baths (62 pages)
- ME Heating Immersion Circulator (61 pages)
- MB-17 Open Heating Bath Circulators (52 pages)
- MB Heating Immersion Circulator (48 pages)
- Magnum 91 Temperature System (77 pages)
- MA-4 Heating Circulators (51 pages)
- MA Heating Immersion Circulator (50 pages)
- LH 47 Highly Dynamic Temperature Control Systems (76 pages)