8.5. Analog

inputs/outputs








or


This submenu enables setting of the input and output values
for the programmer input and the temperature recorder outputs
of socket REG+E-PROG (21).

 Press enter

to select the inputs/outputs submenu.

 Use the up/down cursor keys

to select the

desired option and press enter

to open.

Chan.1

voltage output for recorder (V)

Chan.2

voltage output for recorder (V)

Chan.3

current output for recorder

(mA)

EPROG

external programmer input

First define the desired output value for channels 1 to 3:

Press the P-key

to select the desired output value and set

with enter


Setpoint

active setpoint temperature

(T1, T2, integr. programmer/ext. programmer)

ActInt internal actual temperature value
(bath

temperature)

ActExt external actual temperature value
(external

sensor)

Power

periodic or intermittent heating or cooling

 Then select the display size for channels 1 to 3:


Channel 1 and 2 voltage outputs
Assign the voltage values of 0 V to the lowest and 10 V to
the highest necessary temperature or power rating required
as an output value (°C or %).
Current output channel 3
Assign the current values 0 mA or 4 mA to the lowest and
20 mA to the hightest temperature or power rating required
as an output value (°C or %).
The current output offers 2 ranges for selection:
0 to 20 mA and 4 to 20 mA.

Select the desired range by pressing the P-key

and set with enter

.

The LCD display changes automatically.