JULABO FC600-T Recirculating cooler User Manual
Page 44

Menu functions
44
ACTUAL
SET
Set the external programmer to the lowest temperature
value.
In the calibration level, select the item AdP and set
the lowest temperature value (example: -20 °C).
Press the setpoint key
(E).
Follow the instructions under section 6.3.1.
page 27.
Do not alter any setting on either of the units as long as
the display "Ad_P" is blinking.
The lowest programmer value of -20 °C, and the
highest value of 80 °C correspond now to the
respective values of the recirculating cooler.
Example:
Programmer
Recirculating
cooler
-99.9 °C
-20
°C
- 50 °C
-10
°C
0
°C
0
°C
50
°C 10
°C
100
°C 20
°C
200
°C 40
°C
300
°C 60
°C
400 °C
80
°C
This manual is related to the following products:
- FC1200-T Recirculating cooler FC1600-T Recirculating cooler FCW2500T Recirculating cooler FCW1600T Recirculating cooler FCW1200T Recirculating cooler FCW1600S Recirculating cooler FCW1200S Recirculating cooler FC1600T Recirculating cooler FC1200T Recirculating cooler FC1600S Recirculating cooler FC1200S Recirculating cooler