JULABO FC600-T Recirculating cooler User Manual
Page 39

Recirculating coolers
39
9.3. Status
messages
Message
Description - Recirculating cooler ...
00 MANUAL STOP
... in condition "OFF" (LOCAL)
01 MANUAL START
... in keypad control mode (LOCAL)
02 REMOTE STOP
... in condition "rOFF" (RS 232)
04 REMOTE START
... in remote control mode (RS 232)
9.4. Error
messages
Message Description
-01 WORKING PROCESSOR ALARM
Internal error
-02 EXCESS TEMPERATURE ALARM
High temperature alarm
-03 LOW TEMPERATURE ALARM
Low temperature alarm
-04 SENSOR DIFFERENCE ALARM
Sensor difference alarm
Safety sensor
Feed
C
25
-05 LOW LEVEL ALARM
Low liquid level alarm
-06 PROCESSOR COMMUNICATION
ERROR
Internal error
-07 I2C-BUS WRITE ERROR
Internal error
-08 I2C-BUS READ ERROR
Internal error
-09 I2C-BUS READ/WRITE ERROR
Internal error
-10 COMMAND NOT ALLOWED IN
CURRENT OPERATING MODE
Invalid command in current operating mode
-11 EXTERNAL SENSOR ALARM
Error on external Pt100 sensor
-12 VALUE TOO SMALL
Value too small
-13 VALUE TOO LARGE
Value too large
-14 INVALID COMMAND
Invalid command
-15 WARNING: STAND-BY PLUG IS
MISSING
External stand-by plug is missing
(see page 36)
-16 WARNING: VALUE EXCEEDS
TEMPERATURE LIMITS
Value lies outside the permissible range for
the safety temperature limits. But value is
stored anyway.
-17 PUMP ERROR
Pump disconnection
-18 COMPRESSOR ERROR
Compressor disconnection
-19 HEATER TRIAC SHORTED
Heater disconnection
- FC1200-T Recirculating cooler FC1600-T Recirculating cooler FCW2500T Recirculating cooler FCW1600T Recirculating cooler FCW1200T Recirculating cooler FCW1600S Recirculating cooler FCW1200S Recirculating cooler FC1600T Recirculating cooler FC1200T Recirculating cooler FC1600S Recirculating cooler FC1200S Recirculating cooler