JULABO FC600-T Recirculating cooler User Manual
Page 36
Electrical connections
36
Stand-by input (L)
1
2
3
AK
Pin
assignment:
Pin
Signal
1
not
used
2
5 V / DC
3
0 V
Activate the stand-by input:
Under menu item E_Sb, set the parameter to 1
(see page 41).
Connect an external contact 'AK' (e.g. for emergency switch-off)
or an alarm contact of the superordinated application system.
In case the connection between Pin 2 and Pin 3 is interrupted by
opening the contact 'AK', a complete shutdown of the circulating
pump, heater and cooling compressor is effected, and the unit
enters the condition "stand-by". The message "E_Sb" appears on
the MULTI-Display (LED2) (see page 34).
Return flow safety device
Control connector for solenoid valves (14.)
(line voltage: 230 V).
This manual is related to the following products:
- FC1200-T Recirculating cooler FC1600-T Recirculating cooler FCW2500T Recirculating cooler FCW1600T Recirculating cooler FCW1200T Recirculating cooler FCW1600S Recirculating cooler FCW1200S Recirculating cooler FC1600T Recirculating cooler FC1200T Recirculating cooler FC1600S Recirculating cooler FC1200S Recirculating cooler