JULABO FC600-T Recirculating cooler User Manual

Page 38

background image

Remote control

38

Command

Parameter

Response of recirculating cooler

out_sp_00
in_sp_00

xx.x
no

Set working temperature value
Ask for working temperature value

in_sp_01

no

Ask for high temperature value

in_sp_02

no

Ask for low temperature value

out_sp_03

in_sp_03

xxx

no

Set control ratio for feed/return flow temperature
Ask for actual control ratio

out_sp_07
in_sp_07

xx.xx
xx.xx

Maximum value of the internal temperature in case of
external control (HL - High Limit)
out = set; in = ask

out_sp_08
in_sp_08

xx.xx
xx.xx

Minimum of the internal temperature in case of external
control (LL - Low Limit)
out = set; in = ask

in_pv_00

no

Ask for actual feed temperature

in_pv_01

no

Ask for actual temperature of external sensor

in_pv_02

no

Ask for actual heater capacity

in_pv_03

no

Ask for actual return temperature

in_pv_04

no

Ask for actual safety temperature

out_par_06
in_par_06

x:x
x:x

Control parameter Xp of the internal controller
out = set; in = ask

out_par_07
in_par_07

xx.xx
xx.xx

Control parameter Tn of the internal controller
out = set; in = ask

out_par_08
in_par_08

xx.xx
xx.xx

Control parameter Tv of the internal controller
out = set; in = ask

out_par_09
in_par_09

xx.xx
xx.xx

Control parameter Xp of the external controller
out = set; in = ask

out_par_10
in_par_10

xx.xx
xx.xx

Control parameter Xpu of the cascaded P-controller in case
of external control
out = set; in = ask

out_par_11
in_par_11

xx.xx
xx.xx

Control parameter Tn of the external controller
out = set; in = ask

out_par_12
in_par_12

xx.xx
xx.xx

Control parameter Tv of the external controller
out = set, in = ask