Attention:
The following menu point Cont is only adjustable via the
configuration level if the status is „OFF“.
On a unit which is switched on this status is reached by
switching on/off at the mains switch (1).

ACTUAL

SET

Cont - Control mode
0

=

internal

control

1

= external control (with an external

sensor connected to "N")






10.4. Selecting/exiting

the calibration level

+ 2x

+ 2x

Simultaneously press the
"MENUE" key (4.2) and twice

the edit key " " to select the calibration level
or

the edit key " " to exit the calibration level.

+

Select the menu items of the calibration level one by one
by pressing simultaneously the menu key and one of the
cursors.


10.5. Calibration

ACTUAL

SET

Ad E - Calibration of external sensor

Connect a Pt100 sensor to the connector (N), immerse

into the calibration bath. Use a calibration thermometer
and measure the bath temperature.
Adjust the external sensor to this value (example: 20.8 °C).

Press the setpoint key

(E).

Follow the instructions under section 6.3.1. page 27