beautypg.com

JULABO FC600-T Recirculating cooler User Manual

Page 29

background image

Recirculating coolers

29

6.3.3. Setting the safety temperatures



F G

This safety function is independent of the control circuit.

Press the desired setpoint key (F, G).

Follow the instructions under section 6.3.1.
page 27



Recommendation:

Set the high temperature limit at least 5 K above the actual
bath temperature.

Set the low temperature limit at least 5 K below the setpoint.



+

ACTUAL

SET

When the temperature of the bath liquid reaches the limits of
the safety values, a complete shutdown of the circulating
pump, heater and cooling compressor is effected.
The alarm light illuminates and an audible signal is
triggered. An error message appears on the MULTI-
DISPLAY (LED 2)
(see page 33).



Turn the mains switch (1.) off and on. The alarm state is
cancelled and the circulator is put back into operation.
(Switching on - see page 25).

6.3.4. Low liquid level protection


+

ACTUAL

SET

As soon as the bath liquid falls below the "MIN" level
(10.), a complete shutdown of the circulating pump,
heater, and cooling compressor is effected.
The alarm light illuminates and an audible signal is
triggered. An error message appears on the MULTI-
DISPLAY (LED 2)
(see page 33).