Manual operation

32


Optimization instructions for the PID control
parameters:

The heat-up curve reveals inappropriate control settings


optimum setting

Inappropriate settings may produce the following heat-up
curves:


Xp too low


Tv/Tn too low


Xp
or
Tv too high